An integrated beamforming driver for CMUT based ultrasound catheter ablation system

Yetik, Hasan and Ergün, Arif Sanli and Bozkurt, Ayhan (2014) An integrated beamforming driver for CMUT based ultrasound catheter ablation system. In: IEEE International Ultrasonics Symposium (IUS 2014), Chicago, IL

Full text not available from this repository. (Request a copy)

Abstract

In CMUT based ultrasound catheter ablation systems, transducer arrays and their drivers are to be kept as small as possible due to size restrictions, while maintaining a reasonable number of channels for proper beamforming. In addition, these systems should be highly integrated with minimal power consumption for flexibility and functionality. Despite the size restrictions on these systems, such problems can be overcome by ASICs. In this work, we present a simple 8 channel high-frequency, high-voltage driver, together with its beamforming and control circuitry designed with AMS's H35 high voltage process.
Item Type: Papers in Conference Proceedings
Subjects: T Technology > TK Electrical engineering. Electronics Nuclear engineering
Divisions: Faculty of Engineering and Natural Sciences > Academic programs > Electronics
Faculty of Engineering and Natural Sciences