A new nanocrystalline diamond-based biosensor for the detection of cardiovascular risk markers

Warning The system is temporarily closed to updates for reporting purpose.

Qureshi, Anjum and Kallempudi, S. Saravan and Gürbüz, Yaşar and Howell, Mick and Kang, Weng P. and Davidson, Jimmy L. (2009) A new nanocrystalline diamond-based biosensor for the detection of cardiovascular risk markers. In: 23rd Eurosensors Conference, Lausanne, Switzerland

[thumbnail of This is a RoMEO green publisher -- author can archive post-print (ie final draft post-refereeing)] PDF (This is a RoMEO green publisher -- author can archive post-print (ie final draft post-refereeing))
A_new_nanocrystalline_diamond-based_biosensor_for_the_detection_of_cardiovascular_risk_markers.pdf

Download (265kB)

Abstract

In this paper, a new method to probe associative interactions of C-reactive protein (CRP) antigen with CRP antibody immobilized on a gold-interdigitated diamond electrodes was investigated. The CRP antigen detection was performed by capacitive/dielectric-constant measurements. Our results showed that the dynamic detection range using optimized conditions for a given antibody concentration (100 μg/ml) was found to be in the range 25-800 ng/ml of CRP-antigen. Biosensor developed in this study can be potentially used for detection of elevated CRP levels in suspected subjects for early diagnosis.
Item Type: Papers in Conference Proceedings
Uncontrolled Keywords: Nanocrystalline diamond; Gold interdigitated; C-reactive protein; Biosensor
Divisions: Faculty of Engineering and Natural Sciences > Academic programs > Electronics
Faculty of Engineering and Natural Sciences
Depositing User: Yaşar Gürbüz
Date Deposited: 30 Dec 2009 16:14
Last Modified: 26 Apr 2022 08:54
URI: https://research.sabanciuniv.edu/id/eprint/13640